A law enforcement grade breathalyzer
A
breathalyzer (a
portmanteau of
breath and
analyzer) is a device for estimating
blood alcohol content (BAC) from a
breath sample. "Breathalyzer" is the brand name of a series of
models made by one manufacturer of these instruments (originally
Smith and Wesson, later it was sold
to
National Draeger), but has become a
genericized trademark for all
such instruments. In Canada, a preliminary non-evidentiary
screening device can be approved by Parliament as an
approved screening device and an
evidentiary breath instrument can be similarly designated as an
approved instrument. The U.S.
National Highway Traffic Safety Administration maintains a
"Conforming Products List" of breath alcohol devices approved for
evidentiary use, as well as for preliminary screening use.
Origins
A 1927 paper produced by Emil Bogen who collected air in a football
and then tested this air for traces of alcohol discovered that the
alcohol content of 2 litres of expired air was a little greater
than that of 1cc of urine. However, research into the possibilities
of using breath to test for alcohol in a person's body dates as far
back as 1874 when Anstie made the observation that small amounts of
alcohol were excreted in breath.
The first practical roadside breath testing device intended for use
by the police was the drunkometer. The drunkometer was developed by
Professor Harger in 1938. The drunkometer collected a motorist's
breath sample directly into a balloon inside the machine. The
breath sample was then pumped through an acidified
potassium permanganate solution. If
there was alcohol in the breath sample, the solution changed
colour. The greater the colour change, the more alcohol there was
present in the breath.
The drunkometer was quite cumbersome and was approximately the size
of a shoe box. It was more reminiscent of a portable
laboratory.
In late
1927, in a case in Marlborough
, England, a Dr. Gorsky, Police Surgeon, asked a
suspect to inflate a football bladder with his breath. Since
the 2 liters of the man's breath contained 1.5 ml of ethanol, Dr.
Gorsky testified before the court that the defendant was "50%
drunk".Though technologies for detecting alcohol vary, it's widely
accepted that Dr. Robert Borkenstein (1912–2002), a captain with
the Indiana State Police and later a professor at Indiana
University at Bloomington, is regarded as the first to create a
device that measures a subject's blood alcohol level based on a
breath sample. In 1954, Borkenstein invented his breathalyzer,
which used chemical oxidation and photometry to determine alcohol
concentration. Subsequent breathalyzers have converted primarily to
infrared spectroscopy. The
invention of the breathalyzer provided law enforcement with a
non-invasive test providing immediate results to determine an
individual's breath alcohol concentration at the time of testing.
Also, the breath alcohol concentration test result itself can vary
between individuals consuming identical amounts of alcohol due to
gender, weight, and genetic pre-disposition.
Chemistry
When the user exhales into the breathalyzer, any ethyl alcohol
present in their breath is
oxidized to
acetic acid at the
anode:
CH
3CH
2OH(g) + 4OH
-(aq) ->
HC
2H
3O
2(g) + 3H
2O(l) +
4e
-
At the
cathode, oxygen gas from the
atmosphere is
reduced:
O
2(g) + 2H
2O(l) + 4e
- ->
4OH
-(aq)
The overall reaction, then, is the oxidation of ethyl alcohol to
acetic acid and water.
CH
3CH
2OH(g) + O
2(g) ->
HC
2H
3O
2(g) +
H
2O(l)
The
electrical current produced
by this reaction is measured, processed, and displayed as an
approximation of overall blood alcohol content by the
breathalyzer.
The
hydroxide ion is provided by aqueous
Potassium Hydroxide or Sodium Hydroxide, depending on the
manufacturer.
Law enforcement
Breath analyzers do not directly measure blood alcohol content or
concentration, which requires the analysis of a blood sample.
Instead, they estimate BAC indirectly by measuring the amount of
alcohol in one's
breath. Two form factors are most prevalent. Desktop
analyzers generally utilize infrared
spectrophotometer technology,
electrochemical
fuel cell technology, or a
combination of the two. Hand-held field testing devices, are
generally based on electrochemical
fuel
cell analysis, and depending upon jurisdiction may be used by
officers in the field as a form of "field sobriety test" commonly
called PBT (preliminary breath test) or PAS (preliminary alcohol
screening), or as evidential devices in POA (point of arrest)
testing.
Consumer use
There are many models of consumer or personal breath alcohol
testers on the market. These hand-held devices are generally less
expensive than the devices used by law enforcement. Many consumer
breath testers use semiconductor based sensing technology which is
both less expensive and less accurate and reliable than fuel cell
or infrared devices. While these devices can sense the presence of
breath alcohol they do not provide reliable measures of BAC.
Semiconductor devices are also more prone to false positives, that
is they can signal alcohol even when none is present.
More recently manufacturers have begun to introduce consumer breath
testers that use the same platinum fuel cell technology as used in
law enforcement portable breath testers. Some public breath alcohol
testers such as
Alcospot, use fuel cell
sensors as well, which encourage their high accurancy and long
durability. The LifeGuard breathtester from Lifeloc Technologies is
an example of a personal breath tester that uses the same fuel cell
technology that Lifeloc uses in their FC 10 law enforcement
testers. These devices are more expensive than semiconductor
models, however they are also more precise, reliable and accurate.
Another advantage of all fuel cell devices is that they are less
prone to false positives than are semiconductor devices.
All breath alcohol testers sold to consumers in the United States
are required to be certified by the
Food and Drug Administration,
while those used by law enforcement must be approved by the
Department of Transportation's
National Highway
Traffic Safety Administration.
Manufacturers of over the counter consumer breathalyzers must
submit a FDA 510(k) Premarket Clearance to demonstrate that the
device to be marketed is at least as safe and effective, that is,
substantially equivalent, to a legally marketed device (21 CFR
807.92(a) (3)) that is not subject to Premarket Approval (PMA).
Submitters must compare their device to one or more similar legally
marketed devices and make and support their substantial equivalency
claims.
Breath test evidence in the United States
The breath alcohol content reading is used in criminal prosecutions
in two ways. The operator of a vehicle whose reading indicates a
BrAC over the legal limit for driving will be charged with having
committed an
illegal per se
offense: that is, it is a misdemeanor throughout the United States
to drive a vehicle with a BrAC of 0.08 or higher (0.00-0.05 in all
states for drivers under 21). One exception is the State of
Wisconsin, where a first time drunk driving offense is normally a
civil ordinance violation. The breathalyzer reading will be offered
as evidence of that crime, although the issue is what the BrAC was
at the time of driving rather than at the time of the test. Some
jurisdictions now allow the use of breathalyzer test results
without regard as to how much time passed between operation of the
vehicle and the time the test was administered. The suspect will
also be charged with
driving under the influence of
alcohol (sometimes referred to as driving or operating while
intoxicated). While BrAC tests are not necessary to prove a
defendant was under the influence, laws in most states require the
jury to presume that he was under the influence if his BrAC if
found and believed to be over 0.08 (grams of alcohol/210 liters
breath), when driving. This is a
rebuttable presumption, however: the
jury can disregard the test if they find it unreliable or if other
evidence establishes a reasonable doubt.
If a defendant refuses to take a breathalyzer test, most states
allow evidence of that fact to be introduced; in many states, the
jury is instructed that they can draw a permissible inference of
"consciousness of guilt." Many states also operate under "implied
consent," meaning that anyone issued a driver's license in the
state agrees to submit to a test of his or her breath, blood, or
urine when requested by a law enforcement officer. Failure to
submit to such a test may result in automatic suspension of his or
her driver's license even if not convicted of drunk driving.
Failure to submit to such a test may also serve to enhance the
penalties for a drunk driving conviction. In drunk driving cases in
Massachusetts and Delaware, if the defendant refuses the
breathalyzer there can be no mention of the test during the
trial.
In Massachusetts, a suspect who refused a BAC test will have
his/her driver's license will be administratively suspended for a
period of time (time period details needed), however, the jury in a
DUI case may not be told of the suspect's refusal to take the test.
The model jury instructions require that the judge tell the jury
there are many possible reasons for the absence of a BAC test, that
they should make their decision based on other available info, and
draw no inference for either side from the absence of such a test.
Suspects who refuse the test face license suspension and legal
fees, but have a much greater chance of beating the charge than
suspects who take the test.
Instruments such as the
Intoxilyzer
5000 by
CMI, Inc. are known as
"evidentiary breath testers" and generally produce court-admissible
results. Other instruments, such as the
Intoxilyzer SD-2, also by CMI, Inc., or the
Alcosensor III by
Intoximeters or the
FC10 by
Lifeloc Technoloies, are known
as "preliminary breath testers" (PBT), and their results, while
valuable to an officer attempting to establish probable cause for a
drunk driving arrest, are generally not admissible in court. Some
states do not permit data or "readings" from hand-held PBTs to be
presented as evidence in court. They are generally admissible, if
at all, only to show the presence of alcohol or as a pass-fail
field sobriety test to help
determine probable cause to arrest.
South Dakota
does not permit data from any type or size of
breath tester but relies entirely on blood tests to ensure
accuracy.
Common sources of error
Breath testers can be very sensitive to temperature, for example,
and will give false readings if not adjusted or recalibrated to
account for ambient or surrounding air temperatures. The
temperature of the subject is also very important.
Breathing pattern can also significantly affect breath test
results. One study found that the BAC readings of subjects
decreased 11 to 14% after running up one flight of stairs and
22–25% after doing so twice. Another study found a 15% decrease in
BAC readings after vigorous exercise or hyperventilation.
Hyperventilation for 20 seconds has been
shown to lower the reading by approximately 32%. On the other hand,
holding one's breath for 30 seconds can increase the breath test
result by about 28%.
Some breath analysis machines assume a
hematocrit (cell volume of blood) of 47%.
However, hematocrit values range from 42 to 52% in men and from 37
to 47% in women. A person with a lower hematocrit will have a
falsely high BAC reading.
Failure of law enforcement officers to use the devices properly or
of administrators to have the machines properly maintained and
re-calibrated as required are particularly common sources of error.
However, most states have very strict guidelines regarding officer
training and instrument maintenance and calibration.
Research indicates that breath tests can vary at least 15% from
actual blood alcohol concentration. An estimated 23% of individuals
tested will have a BAC reading higher than their true BAC.
Police in
Victoria,
Australia
use breathalyzers that give a recognized 20 percent
tolerance on readings. Noel Ashby, former Victoria Police
Assistant Commissioner (Traffic & Transport), claims that this
tolerance is to allow for different body types.
Calibration
Most handheld breathalyzers sold to consumers use a silicon oxide
sensor to determine the blood alcohol concentration. Most
professional use breath alcohol testers use fuel cell sensors.
Without calibration, the accuracy of these sensors degrades over
time and with repeated use. The calibration process aims to focus
the sensor's ability to detect an accurate reading. New advances in
breathalyzer design allow some models to self-calibrate or easily
replace the sensor module without the need to send the unit to a
calibration lab.
Non-specific analysis
One major problem with older breathalyzers is non-specificity: the
machines not only identify the
ethyl
alcohol (or
ethanol) found in alcoholic
beverages, but also other substances similar in molecular structure
or reactivity.
The oldest breathalyzer models pass breath through a solution of
potassium dichromate, which
oxidizes ethanol into
acetic acid, changing color in the process. A
monochromatic light beam is passed
through this sample, and a detector records the change in intensity
and, hence, the change in color, which is used to calculate the
percent alcohol in the breath. However, since potassium dichromate
is a strong oxidizer, numerous alcohol groups can be oxidized by
it, producing
false positives. This
source of false positives is unlikely as very few other substances
found in exhaled air is oxidisable.
Infrared-based breathalyzers project an infrared beam of radiation
through the captured breath in the sample chamber and detect the
absorbance of the compound as a function
of the
wavelength of the beam, producing
an absorbance spectrum that can be used to identify the compound,
as the absorbance is due to the
harmonic vibration and stretching of
specific bonds in the molecule at specific wavelengths (see
infrared spectroscopy). The
characteristic bond of alcohols in infrared is the O-H bond, which
gives a strong absorbance at a short wavelength. The more light is
absorbed by compounds containing the
alcohol
group, the less reaches the detector on the other side—and the
higher the reading. Other groups, most notably
aromatic rings and
carboxylic acids can give similar
absorbance readings. Even water vapour does.
Interfering compounds
Some natural and volatile interfering compounds do exist, however.
For example, the
National Highway
Traffic Safety Administration (NHTSA) has found that dieters
and
diabetics may have
acetone levels hundreds and even thousand of times
higher than those in others. Acetone is one of the many substances
that can be falsely identified as ethyl alcohol by some breath
machines. However, new machines like the Draeger Breathalyzer use
technology that filters out substances like acetone.
A study in Spain showed that metered-dose inhalers (MDIs) used in
asthma treatment are also a cause of false positives in breath
machines.
Substances in the environment can also lead to false BAC readings.
For example, methyl tert-butyl ether (
MTBE), a
common gasoline additive, has been alleged anecdotally to cause
false positives in persons exposed to it. Tests have shown this to
be true for older machines; however, newer machines detect this
interference and compensate for it. Any number of other products
found in the environment or workplace can also cause erroneous BAC
results. These include compounds found in
lacquer, paint remover,
celluloid,
gasoline, and
cleaning fluids, especially
ethers,
alcohols, and other
volatile compounds.
Homeostatic variables
Breathalyzers assume that the subject being tested has a 2100-to-1
"partition ratio" in converting alcohol measured in the breath to
estimates of alcohol in the blood. If the instrument estimates the
BAC, then it measures weight of alcohol to volume of breath, so it
will effectively measure grams of alcohol per 2100 ml of breath
given. This measure is in direct proportion to the amount of grams
of alcohol to every 100 ml of blood. Therefore, there is a 2100 to
1 ratio of alcohol in blood to alcohol in breath. However, this
assumed "partition ratio" varies from 1300:1 to 3100:1 or wider
among individuals and within a given individual over time. Assuming
a true (and US legal) blood-alcohol concentration of .07%, for
example, a person with a partition ratio of 1500:1 would have a
breath test reading of .10%—over the legal limit.
Most individuals do, in fact, have a 2100-to-1 partition ratio in
accordance with William
Henry's law,
which states that when the water solution of a volatile compound is
brought into equilibrium with air, there is a fixed ratio between
the concentration of the compound in air and its concentration in
water. This ratio is constant at a given temperature. The human
body is 37 degrees Celsius on average. Breath leaves the mouth at a
temperature of 34 degrees Celsius. Alcohol in the body obeys
Henry's Law as it is a volatile compound and diffuses in body
water. To ensure that variables such as fever and hypothermia could
not be pointed out to influence the results in a way that was
harmful to the accused, the instrument is calibrated at a ratio of
2100:1, underestimating by 9 percent. In order for a person running
a fever to significantly overestimate, he would have to have a
fever that would likely see the subject be in the hospital rather
than driving in the first place. Studies suggest that about 1.8% of
the population have a partition ratio below 2100. Thus, a machine
using a 2100-to-1 ratio could actually under-report. As much as 14%
of the population has a partition ratio above 2100, thus causing
the machine to overestimate the BAC.
Further, the assumption that the test subject's
partition
ratio will be average—that there will be 2100 parts in the
blood for every part in the breath—means that accurate analysis of
a given individual's blood alcohol by measuring breath alcohol is
difficult, as the ratio varies considerably.
Variance in how much one breathes out can also give false readings,
usually low. This is due to biological variance in breath alcohol
concentration as a function of the volume of air in the lungs, an
example of a factor which interferes with the liquid-gas
equilibrium assumed by the devices. The presence of volatile
components is another example of this; mixtures of volatile
compounds can be more volatile than their components, which can
create artificially high levels of ethanol (or other) vapors
relative to the normal biological blood/breath alcohol
equilibrium.
Mouth alcohol
One of the most common causes of falsely high breathalyzer readings
is the existence of
mouth alcohol. In analyzing a
subject's breath sample, the breathalyzer's internal computer is
making the assumption that the alcohol in the breath sample came
from
alveolar air—that is, air exhaled from deep within
the lungs. However, alcohol may have come from the mouth, throat or
stomach for a number of reasons. To help guard against
mouth-alcohol contamination, certified breath test operators are
trained to carefully observe a test subject for at least 15–20
minutes before administering the test.
The problem with mouth alcohol being analyzed by the breathalyzer
is that it was not absorbed through the stomach and intestines and
passed through the blood to the lungs. In other words, the
machine's computer is mistakenly applying the "partition ratio"
(see above) and multiplying the result. Consequently, a very tiny
amount of alcohol from the mouth, throat or stomach can have a
significant impact on the breath alcohol reading.
Other than recent drinking, the most common source of mouth alcohol
is from belching or
burping. This causes the
liquids and/or gases from the stomach—including any alcohol—to rise
up into the soft tissue of the
esophagus
and oral cavity, where it will stay until it has dissipated. The
American Medical Association concludes in its Manual for Chemical
Tests for Intoxication (1959): "True reactions with alcohol in
expired breath from sources other than the alveolar air
(eructation, regurgitation, vomiting) will, of course, vitiate the
breath alcohol results." For this reason, police officers are
supposed to keep a DUI suspect under observation for at least 15
minutes prior to administering a breath test. Instruments such as
the Intoxilyzer 5000 also feature a "slope" parameter. This
parameter detects any decrease in alcohol concentration of .006 g
per 210L of breath in 6/10th's of a second, a condition indicative
of residual mouth alcohol, and will result in an "invalid sample"
warning to the operator, notifying the operator of the presence of
the residual mouth alcohol. PBT's, however, feature no such
safeguard.
Acid reflux, or
gastroesophageal reflux
disease, can greatly exacerbate the mouth alcohol problem. The
stomach is normally separated from the throat by a valve, but when
this valve becomes herniated, there is nothing to stop the liquid
contents in the stomach from rising and permeating the esophagus
and mouth. The contents—including any alcohol—are then later
exhaled into the breathalyzer.
Mouth alcohol can also be created in other ways. Dentures, for
example, will trap alcohol. Periodental disease can also create
pockets in the gums which will contain the alcohol for longer
periods. Also known to produce false results due to residual
alcohol in the mouth is passionate kissing with an intoxicated
person. And recent use of mouthwash or breath freshener—possibly to
disguise the smell of alcohol when being pulled over by
police—contain fairly high levels of alcohol.
Testing during absorptive phase
Absorption of alcohol continues for anywhere from 20 minutes (on an
empty stomach) to two and one half hours (on a full stomach) after
the last consumption. Peak absorption generally occurs within an
hour. During the initial absorptive phase, the distribution of
alcohol throughout the body is not uniform. Uniformity of
distribution; called equilibrium; occurs just as absorption
completes. In other words, some parts of the body will have a
higher blood alcohol content (BAC) than others. One aspect of the
non-uniformity before absorption is complete is that the BAC in
arterial blood will be higher than in
venous blood. Laws generally require blood
samples to be venous.
During the initial absorption phase, arterial blood alcohol
concentrations are higher than venous. After absorption, venous
blood is higher. This is especially true with bolous dosing. With
additional doses of alcohol the body can reach a sustained
equilibrium, when absorption and elimination are proportional.
Calculating a
general absorption rate of
0.02/drink and a
general elimination rate of
0.015/hour. (One drink = 1.5 ounce liquor/12 ounce beer/5 ounce
wine
[117693].)
Breath alcohol is a representation of the equilibrium of alcohol
concentration as the blood gases (alcohol) pass from the (arterial)
blood into the lungs to be expired in the breath. And the venous
blood picks up oxygen for distribution throughout the body. Breath
alcohol concentrations are generally lower than blood alcohol
concentrations because a true representation of blood alcohol
concentration is only possible if the lungs were able to completely
deflate. Vitreous (eye) fluid provides the most accurate account of
blood alcohol concentrations.
Retrograde extrapolation
The breathalyzer test is usually administered at a police station,
commonly an hour or more after the arrest. Although this gives the
BrAC at the time of the test, it does not by itself answer the
question of what it was at the time of driving. The prosecution
typically provides an estimated alcohol concentration at the time
of driving utilizing retrograde extrapolation, presented by expert
opinion. This involves projecting back in time to estimate the BrAC
level at the time of driving, by applying the physiological
properties of absorption and elimination rates in the human
body.Extrapolation is calculated using five factors and a general
elimination rate of 0.015/hour.For example: Time of breath
test-10:00pm...Result of breath test-0.080...Time of driving-9:00pm
(stopped by officer)...Time of last drink-8:00pm...Last
food-12:00pmUsing these facts an expert can say the person's last
drink was consumed on an empty stomach. Which means absorption of
the last drink (at 8:00) was complete within one hour-9:00. At the
time of the stop the driver is fully absorbed. The test result of
0.080 at 10:00. So we add in the one hour of elimination that has
occurred since the stop. 0.080+0.015=0.095 approximate breath
alcohol concentration at the time of the stop.
Photovoltaic assay
The photovoltaic assay, used only in the dated
Intoximeter 3000, is a form of breath
testing rarely encountered today. The process works by using
photocells to analyze the color change of
a
redox (oxidation-reduction) reaction. A
breath sample is bubbled through an
aqueous
solution of
sulfuric acid,
potassium dichromate, and
silver nitrate. The
silver nitrate acts as a
catalyst, allowing the
alcohol to be oxidized at an appreciable rate. The
requisite acidic condition needed for the reaction might also be
provided by the
sulfuric acid. In
solution,
ethanol
reacts with the
potassium
dichromate, reducing the
dichromate
ion to the
chromium
(III)
ion. This reduction results in a change of
the
solution's colour from red-orange to
green. The reacted
solution is compared to
a
vial of nonreacted
solution by a
photocell,
which creates an
electric current
proportional to the degree of the colour change; this current moves
the needle that indicates BAC.
Like other methods, breath testing devices using
chemical analysis are somewhat prone to
false readings.
Compounds which
have compositions similar to
ethanol, for
example, could also act as reducing agents, creating the necessary
color change to indicate increased BAC.
Breathalyzer myths
There are a number of substances or techniques that can supposedly
"fool" a breathalyzer (i.e. generate a lower
blood alcohol content).
A 2003 episode of the popular science television show
MythBusters tested a number of methods that
supposedly allow a person to fool a breathalyzer test. The methods
tested included
breath mints,
onions,
denture cream,
mouthwash, pennies and batteries; All of
these methods proved ineffective. The show noted that using items
such as
breath mints,
onions,
denture cream and
mouthwash to cover the smell of alcohol
may fool a person but since they will not actually reduce a
person's BAC, there will be no effect on a breathalyzer test
regardless of the quantity used. Pennies supposedly produce a
chemical reaction while batteries supposedly create an electrical
charge, yet neither of these methods affected the breathalyzer
results.
The Mythbusters episode also pointed out another complication
involved in any attempt to fool a breathalyzer by placing something
in one's mouth: It would be necessary to insert the item into one's
mouth, take the breath test, and then possibly remove the item -
all of which would have to be accomplished discreetly enough to
avoid alerting the police officers administering the test (who
would certainly become very suspicious if they noticed that a
person was inserting items into their mouth). It would likely be
very difficult, especially for someone in an intoxicated state, to
be able to accomplish such a feat.
In addition, the show noted that breath tests are often verified
with blood tests (which are far more accurate) and that even if a
person somehow managed to fool a breath test, a blood test would
certainly confirm a person's guilt.
Other substances that might reduce the BAC reading include a bag of
activated charcoal concealed in
the mouth (to absorb alcohol vapor), an oxidizing gas (such as
N
2O, Cl
2, O
3, etc.) which would
fool a fuel cell type detector, or an organic interferent to fool
an infra-red absorption detector. The infra-red absorption detector
is more vulnerable to interference than a laboratory instrument
measuring a continuous absorption spectrum since it only makes
measurements at particular discrete wavelengths. However, due to
the fact that any interference can only cause higher absorption,
not lower, the estimated
blood
alcohol content will be overestimated.
A 2007 episode on the
Spike network's show
Manswers, showed some of the more common and
not so common ways of attempts to beat the breathalyzer, none of
which work. Test 1 one was to suck on a copper coin (Actually
copper coins are now generally often only copper-coated and mostly
zinc or steel). Test 2 was to hold a battery on the tongue. Test 3
was to chew gum. None of these tests showed a "pass" reading if the
subject had consumed alcohol.
Products that interfere with testing
On the other hand, products such as mouthwash or
breath spray can "fool" breath machines by
significantly raising test results. Listerine, for example,
contains 27% alcohol; because the breath machine will assume the
alcohol is coming from alcohol in the blood diffusing into the lung
rather than directly from the mouth, it will apply a "partition
ratio" of 2100:1 in computing blood alcohol concentration—resulting
in a false high test reading. To counter this, officers are not
supposed to administer a PBT for 15 minutes after the subject eats,
vomits, or puts anything in their mouth. In addition, most
instruments require that the individual be tested twice at least
two minutes apart. Mouthwash or other mouth alcohol will have
somewhat dissipated after two minutes and cause the second reading
to disagree with the first, requiring a retest. (Also see the
discussion of the "slope parameter" of the Intoxilyzer 5000 in the
"Mouth Alcohol" section above.)
This was clearly illustrated in a study conducted with Listerine
mouthwash on a breath machine and reported in an article entitled
"Field Sobriety Testing:
Intoxilyzers
and Listerine Antiseptic," published in the July 1985 issue of
The Police Chief (p. 70).
Seven individuals were tested at a police station, with readings of
.00%. Each then rinsed his mouth with 20 milliliters of Listerine
mouthwash for 30 seconds in accordance with directions on the
label. All seven were then tested on the machine at intervals of
one, three, five and ten minutes. The results indicated an average
reading of .43 blood-alcohol concentration, indicating a level
that, if accurate, approaches lethal proportions. After three
minutes, the average level was still .020, despite the absence of
any alcohol in the system. Even after five minutes, the average
level was .011.
In another study, reported in 8(22)
Drinking/Driving Law Letter 1, a
scientist tested the effects of Binaca breath spray on an
Intoxilyzer 5000. He performed 23 tests with subjects who sprayed
their throats, and obtained readings as high as .81 — far beyond
lethal levels. The scientist also noted that the effects of the
spray did not fall below detectable levels until after 18
minutes.
References
- State of the Art Breathalyzers: A History
-
http://a257.g.akamaitech.net/7/257/2422/01jan20061800/edocket.access.gpo.gov/2006/pdf/E6-10258.pdf
-
http://www.dot.gov/ost/dapc/testingpubs/20070131_CPL_ASD.pdf
- The Diagnosis of Drunkenness
- Professor Robert F.Borkenstein - An Appreciation of
his Life and Work
- AlcoSpot - Public Alcohol Tester, Technical
info
- FDA > CDRH > Product Classification Database
Search
- Retrieve Pages
- http://www.fda.gov/CDRH/DEVADVICE/314.html
- WI DOT
- Jane Holroyd, Breathalyser's 20 per cent tolerance
defended, Sydney Morning Herald,
16 May 2006
- Organic Chemistry Resources Worldwide
- Elsevier Article Locator
- Forensic-Evidence.com: Biological Evidence/ Breath
Tests for Blood Alcohol Determination: Partition Ratio
- Quantitative measurements of the alcohol
concentra...[Acta Physiol Scand. 1982] - PubMed Result
- Kechagias, et al., "Reliability of Breath-Alcohol Analysis in
Individuals with Gastroesophageal Reflux Disease", 44(4) Journal of
Forensic Sciences 814 (1999).
-
Mythbusters, Season 1, episode 6. First aired November 7,
2003.
External links